  Additional dose assessment from the activation of high-energy linear accelerators used in radiation therapy
 
 
Titel: Additional dose assessment from the activation of high-energy linear accelerators used in radiation therapy
Auteur: Ateia Embarka
Ciraj-Bjelac Olivera
Kovačević Milojko
Beličev Petar
Cvetković Bratislav
Aničin Ivan
Verschenen in: Nuclear technology & radiation protection
Paginering: Jaargang 23 (2008) nr. 2 pagina's 58-64
Jaar: 2008
Inhoud: It is well known that medical linear accelerators generate activation products when operated above certain electron (photon) energies. The aim of the present work is to assess the activation behavior of a medium-energy radiotherapy linear accelerator by applying in situ gamma-ray spectrometry and dose measurements, and to estimate the additional dose to radiotherapy staff on the basis of these results. Spectral analysis was performed parallel to dose rate measurements in the isocenter of the linear accelerator, immediately after the termination of irradiation. The following radioisotopes were detected by spectral analysis: 28Al, 62Cu, 56Mn, 64Cu, 187W, and 57Ni. The short-lived isotopes such as 28Al and 62Cu are the most important factors of the clinical routine, while the contribution to the radiation dose of medium-lived isotopes such as 56Mn, 57Ni, 64Cu, and 187W increases during the working day. Measured dose rates at the isocenter ranged from 2.2 µSv/h to 10 µSv/h in various measuring points of interest for the members of the radiotherapy staff. Within the period of 10 minutes, the dose rate decreased to values of 0.8 µSv/h. According to actual workloads in radiotherapy departments, a realistic exposure scenario was set, resulting in a maximal additional annual whole body dose to the radiotherapy staff of about 3.5 mSv.
Uitgever: VINČA Institute of Nuclear Sciences (provided by DOAJ)
Bronbestand: Elektronische Wetenschappelijke Tijdschriften
 
 

                             Details van artikel 5 van 87 gevonden artikelen
 
<< vorige    volgende >>
 
 Koninklijke Bibliotheek - Nationale Bibliotheek van Nederland