Digital Library
Close Browse articles from a journal
 
<< previous    next >>
     Journal description
       All volumes of the corresponding journal
         All issues of the corresponding volume
           All articles of the corresponding issues
                                       Details for article 3 of 13 found articles
 
 
 
 
 
Title:
Author:
Appeared in: Analytical letters
Paging: Volume 4 (1971) nr. 11 pages 709-714
Year: 1971-11-01
Contents: A gas sampling valve suitable for on-line injection of low pressure (< 0.5 torr) gas mixtures is described. Its advantages over the commercially available valves are enumerated. An error of < 3 per cent due to non-reproducibility in sample injection was observed. A simple arrangement of column and hot wire detector together with this valve enabled quantitative separation of CO and O2 with a lower limit of ∼ 1 × 10-7 moles for each component.
Publisher: Taylor & Francis
Source file: Elektronische Wetenschappelijke Tijdschriften
 
 

                             Details for article 3 of 13 found articles
 
<< previous    next >>
 
 Koninklijke Bibliotheek - National Library of the Netherlands