  Amount of accidental flush by syringe pump due to inappropriate release of occluded intravenous line
 
 
Title: Amount of accidental flush by syringe pump due to inappropriate release of occluded intravenous line
Author: Kawakami, Hiromasa
Miyashita, Tetsuya
Yanaizumi, Ryota
Mihara, Takahiro
Sato, Hitoshi
Kariya, Takayuki
Mizuno, Yusuke
Goto, Takahisa
Appeared in: Technology & health care
Paging: Volume 21 (2013) nr. 6 pages 581-586
Year: 2013-11-13
Contents: BACKGROUND: An unintended bolus is delivered by the syringe pump if intravenous line occlusion is released in an inappropriate manner. OBJECTIVE: The aim of this study was to measure the amount of flushed fluid when an occlusion is inappropriately released and to assess the effect of different syringe pump settings (flow rate, alarm setting, size of syringe and syringe pump model) on the flushed amount. METHODS: After the stopcock was closed, infusions were started with different model syringe pumps (Terufusion® TE312 and TE332S), different syringe sizes or at different alarm settings. After the occlusion alarm sounded, the occlusion was released and the amount of fluid emerging from the stopcock was measured. RESULTS: The bolus was significantly lower when the alarm was set at a low-pressure setting. The bolus was significantly lower with a 10-ml than a 50-ml syringe. A significant difference was seen only when a 50-ml syringe was used (TE312: 1.99 ± 0.16 ml vs. TE332S: 0.674 ± 0.116 ml, alarm High, p < 0.001). CONCLUSION: To minimize the amount of accidentally injected medication, a smaller syringe size and a low alarm setting are important. Using a syringe pump capable of reducing the inadvertently administered bolus may be helpful.
Publisher: IOS Press
Source file: Elektronische Wetenschappelijke Tijdschriften
 
 

                             Details for article 1 of 15 found articles
 
   next >>
 
 Koninklijke Bibliotheek - National Library of the Netherlands