Reliability and validity of a custom-made instrument including a hand-held dynamometer for measuring trunk muscle strength
Titel:
Reliability and validity of a custom-made instrument including a hand-held dynamometer for measuring trunk muscle strength
Auteur:
Jubany, Júlia Busquets, Albert Marina, Michel Cos, Francesc Angulo-Barroso, Rosa
Verschenen in:
Journal of back and musculoskeletal rehabilitation
Paginering:
Jaargang 28 (2014) nr. 2 pagina's 317-326
Jaar:
2014-08-05
Inhoud:
BACKGROUND: Measuring isometric strength is necessary in many areas of health and sport. However, trunk muscles have some particular characteristics that make them difficult to evaluate with simple, inexpensive instruments. OBJECTIVE: To evaluate the reliability and validity of an instrument constructed with a hand-held dynamometer and a metallic structure (HHD+S) for measuring maximum isometric voluntary trunk muscle strength. METHODS: Maximum isometric voluntary trunk muscle strength (extension, flexion and lateral flexion) was measured in 20 healthy individuals using the custom-made instrument (HHD+S) and the gold standard Back-Check (BC). RESULTS: The results showed that the two instruments had high and similar intra-subject reliability. The validity of the HHD+S was demonstrated by the high Pearson coefficient correlation between the two instruments (r ⩾ 0.78). CONCLUSIONS: Given the good trial reliability and the close correlation between the two instruments, we believe that the use of a hand-held dynamometer together with the custom-made metallic structure (HHD+S) allows an evaluation of the maximum isometric voluntary trunk muscle strength to be made, that is very similar in quality, accuracy and reliability to the BC.